The ChemLogic® CLPx Portable Gas Detector provides handheld gas detection to meet a variety of applications. The advanced optics design and automated calibration make the CLPx the most reliable and intuitive gas detector on the market today. A versatile instrument, the CLPx can be configured to detect a variety of gases - eliminating the need to purchase different instruments or gas keys to sample different gases. Only a simple cassette change is required to change from detecting one gas to another.
The CLPx comes in a convenient durable hard-shell storage case with a handle for easy transport. Included in the case are: • A 0-500cc flow meter with a short tube for use when adjusting the pump to verify proper flow at any time during sampling • A battery charger with a charging indication light • A shoulder strap with two rings to attach to the CLPx
The case can also hold a ChemLogic cassette. For replacement cassettes, please click here.
In a bind or want to try-before-you-buy? Check out our short-term loaner program today!
For more information on isocyanates and OSHA standards, please click here. ddd
Lead times vary by item. A shipment/delivery estimate will be provided upon placement of your order. Please contact us to review your delivery requirements before requesting expedited shipping for rush delivery.
Lead times vary by item. A shipment/delivery estimate will be provided upon placement of your order. Please contact us to review your delivery requirements before requesting expedited shipping for rush delivery.
The ChemLogic CLPx is a portable colorimetric gas monitor designed to measure low concentrations of a specific toxic gas or a related range of toxic gases or vapors. It is intended to be used as part of an air quality management program. The monitor uses ChemLogic colorimetric technology, where the tape darkens in the presence of target gas and the instrument’s optics measure that change to determine concentration.
How does the CLPx detect gas?
During analysis, the pump pulls a measured sample of air through the chemically-treated tape in a ChemLogic cassette. If the target gas is present, the tape stains or darkens. The instrument’s optics then measure the difference between the tape’s original color and the stained area. The CLPx converts that optical change into a gas concentration reading and stores the results internally; and, if installed, on the MicroSD card in CSV format.
Is the CLPx a multi-gas monitor?
The CLPx is not a multi-gas monitor in the same sense as a 4-gas personal monitor. It is designed to measure a specific gas or a specific range of related gases/vapors, depending on the gas family and cassette configuration installed at the factory. Gas options available depend on what was purchased and how the unit is configured.
Installation & Power - CLPx Systems
What powers the CLPx?
The CLPx is powered by a 6-volt rechargeable lead-acid battery.
Can I use any charger with the CLPx?
No. You should only use the charger supplied with the CLPx. Using a different charger could damage the battery and/or the instrument.
Is the charger intrinsically safe?
No. The battery charger has not been evaluated for intrinsic safety and must not be used in a potentially explosive atmosphere. Using the charger or CLPx in a flammable or explosive atmosphere could ignite the atmosphere and cause serious injury or death.
Should the CLPx stay on charge when not in use?
Yes, it is recommended to keep the battery at float charge when the unit is not in use. This helps extend battery life and keeps the monitor ready for use.
Can I replace the battery myself?
Yes. To replace the battery, disconnect the charger, make sure the unit is off, work in a clean non-hazardous area, remove the four battery screws with a 9/64 in. (3.5 mm) Allen wrench, disconnect the two-pin connector, and reinstall in reverse order. Battery removal must never be performed in a potentially explosive atmosphere.
How do I power the CLPx on?
To power the unit on, press the Enter button.
Display & Operation - CLPx Systems
What does the asterisk on the analysis screen mean?
An asterisk in the lower-right corner indicates the monitor has passed the time necessary to detect the LDL for the selected gas since the last cassette advance. The asterisk disappears and the timer resets whenever the cassette advances. The operator can also press the down arrow to reset the timer when checking a new location.
What does “SD” on the display mean?
The letters SD appearing on the upper-left of the analysis screen indicate there is an error writing to the MicroSD card or that no MicroSD card is inserted.
What is STEL on the CLPx?
The CLPx automatically calculates STEL for the selected gas. A STEL log is stored in the instrument, and if a properly formatted MicroSD card is installed, STEL details are also written to the card. The monitor keeps at least 16 STEL values depending on memory usage.
What is the “Verify Tape” function for?
The Verify Tape screen shows the currently selected gas and the ChemLogic cassette that should be installed. It is a quick way to confirm that the correct cassette is in the instrument before sampling.
Data Logging & SD Card - CLPx Systems
What does the MicroSD card do?
The MicroSD card stores: Concentration logging, Event history, STEL history, and Configuration information. The SD Card screen also shows remaining free space, this should be checked before use to make sure there is enough room for logging.
Can I remove or insert the MicroSD card while the monitor is running?
No. The manual says to power the machine down first before inserting or removing the MicroSD card, otherwise data may be lost or the card may become corrupted.
What type of MicroSD card format is supported?
Standard MicroSD cards may be used with FAT16 or FAT32 formatting.
How do I install the MicroSD card correctly?
When viewing the unit from the back, install the card with the 8-pin gold edge connector facing to the left.
What should I do if I see “Memory Flash Err”?
The Memory Flash Error can occur due to no SD card installed, SD card full or Card write error. First checks should be whether the MicroSD card is present, has free space, is properly formatted, and is functioning normally.
Alarms & Outputs - CLPx Systems
How do alarms work on the CLPx?
If the concentration exceeds either the low or high alarm setpoint, the monitor will sound an alarm and flash the concentration level. During analysis: Alarm Level 1 blinks at a 2-second rate and Alarm Level 2 blinks at a 1-second rate. The default alarm values are normally set to 1 TLV and 2 TLV when a gas is selected, though they can be adjusted.
Can alarm levels be changed?
Yes. In Setup > Set Alarm Levels, the alarm levels can be adjusted as needed. The manual says to be sure to select Save after making changes.
Can I view past alarm or event history?
Yes. The History screen stores recent events even if there is no SD card installed. The manual says at least 16 events are kept at all times, often more depending on memory use. The most recent event appears first, and you can scroll through events and use the right key to see more details and diagnostic codes.
Sampling & Tubing - CLPx Systems
What is the sampling wand used for?
The sampling wand lets the operator sample gas farther away from the unit, such as around gas cylinders or fittings when leak checking. The sampling wand will come with 5 feet of tubing that can be connected to the inlet.
What is the exhaust fitting option used for?
The exhaust option adds an external exhaust port so sample gas can be routed through tubing to an exhaust hood or similar safe discharge point.
Flow System - CLPx Systems
How do I set the flow?
Go to Setup > Adjust Flow, connect the included flow meter, and adjust the pump until the displayed target flow is reached. A ChemLogic tape cassette must be installed while adjusting flow.
Does flow depend on the gas or cassette?
Yes. The target flow depends on the cassette and gas being used. For example, some gases use 200 cc, while others use 120 cc. That means flow should always be set according to the configured gas/cassette combination rather than assuming one universal value.
Optical System - CLPx Systems
What is the Auto Optics Test?
This is the recommended quick method to verify optics adjustment and light intensity. The manual notes the tape reading will be approximately 2800 during this test.
What should I do if I see optic calibration faults?
Optic calibration faults can occur due to dirty optics, no cassette installed or if you reach the end of the cassette. Verify the correct cassette is installed, check whether the tape is spent, and clean the unit if needed. If the issue remains, contact DOD support.
What is the Cal Stick test?
The Cal Stick Test is not used to calibrate the optics, since the optics self-calibrate, but it does verify that the optics can properly adjust and detect color change. The manual recommends running an Auto Optics Test after using the Cal Stick.
Go to Setup > Select Gas. The available selections are based on the gas family installed at the factory. When you choose a gas, the CLPx loads default alarm levels and gives you the option to go directly to the alarm setting screen.
ChemLogic Cassettes - CLPx Systems
How do I install a ChemLogic cassette?
Open the cassette access door by turning the cassette release knob fully to the left. Remove the black tape guide, place the cassette inside with the label upright and facing outward, and make sure the tape sits properly between the optics block and the inlet nosepiece before closing the door and turning the knob closed.
How long can a cassette be used after opening?
The expiration date of all cassettes is printed on the label of each cassette. However, once the package is opened the maximum number of days the cassette can be used without proper storage is within 45 days. After each use, if the cassette is put back into its original packaging and stored at the proper temperature, it may be used up to the original printed expiration date.
What should I do if I see a tape advance error?
Tape Advance Err1 and Tape Advance Err2 may be caused by a tape jam or low voltage. That means the operator should check battery condition, confirm tape path and cassette seating, and inspect for tape issues or jams.
Software - CLPx Systems
How do I check software version and system information?
Use Setup > System Info. This displays software version, serial number, and selected gas. For version 12.3 and higher, it also shows the installer/updater version.